The Test Equity 140 Temperature Chamber provides full-range temperature testing with the smallest footprint in its class. Based on the field proven technology of the popular Test Equity 1007C, it offers a more compact package at an economical price.
The Test Equity 140 Environmental Chamber features a High/Low Temperature Limit Controller and Alarm. A Limit Controller Provides independent protection against excess temperatures. Both high and low limits can be set. The limit controller will shut down the chamber and trigger the audible alarm in the event of an out of limit condition. A set of uncommitted contacts are available to provide a safety power interlock for your test sample.
Viewing Window: A unique design virtually eliminates exterior window sweating during normal temperature cycling conditions.
Two 4" Access Ports: Allows you to attach wires and sensors to your test sample through both the left and right side of the chamber.
Adjustable Shelf: A heavy-duty stainless steel wire shelf ensures proper airflow around your test sample for uniform temperature distribution.
Chamber Condensate Drain: The chamber floor has a drain to permit removal of condensate from the evaporator that can occur when cycling between low to high temperatures.
Casters: Rugged 3" casters permit easy movement.
Power Cord with Plug: Heavy-duty power cord has a NEMA L6-30P twist-lock plug. Chamber requires 208V or 230V, 1 PH power.
Non-CFC refrigerants are used in a cascade (two compressors) configuration. Expansion valves provide optimum performance throughout the operating range. Pressure regulators keep the operating pressures within safe limits regardless of load conditions. Liquid injection ensures cool compressor operation during high-temperature pull down for long life. Sequential starting of each compressor reduces the current demand on start-up. Each compressor is protected by a pressure switch. Care has been taken to use a minimum number of fittings, joints and welds to reduce the possibility of refrigerant leaks.
All electrical control components are UL approved. Each branch circuit is individually protected with class-CC fuses. The nichrome air-heater has a low watt-density and even heat distribution for reliable performance and fast response. A fusible link provides fail-safe protection against thermal runaways, in addition to the microprocessor-based high/low limit controller.
The side and rear panels are removable to permit complete access to the refrigeration system for ease of service and preventive maintenance.
The 140 chamber can be specified with either the original F4 controller, or optionally, the latest F4T touchscreen controller:
Features and specifications of the Test Equity 140 Chamber include:
Testimonials |
I thought of no place else to try first but you. -- Tony Fremont, CA |