Tektronix Encore




Agilent N8973A, Keysight N8973A, HP N8973A
Noise Figure Analyzer, 10 MHz - 3 GHz

Analyzers >> Phase Noise and Noise Figure Analyzers >> Noise Figure Analyzer, 10 MHz - 3 GHz
Rent Used Test Equipment
Manufacturer: Keysight / Agilent / HP
Model: N8973A

Keysight / Agilent N8973A Noise Figure Analyzer, 10 MHz  - 3 GHz

Electronic Test and Measurement Equipment

Keysight / Agilent N8973A Noise Figure Analyzer, 10 MHz  - 3 GHz Product Datasheet Download Product Datasheet

Rent Test Equipment

Product Overview

The Keysight N8973A (Agilent) Noise Figure Analyzer is a high performance noise figure analyzer designed to make fast, accurate and repeatable noise figure measurement. With the N8973A ease of use features, you can set up complex measurements simply and easily giving you repeatable and reliable results you can trust. The Keysight N8973A offers simultaneous noise figure and gain measurements with the ability to view, print and save the data / display in multiple formats. The N8973A also has the ability to characterize amplifiers and frequency translating devices, including LO control over dedicated 2nd GPIB. The N8973A offers increased measurement accuracy and faster measurement speed over previous Agilent noise figure meters, as well as six user selectable measurement bandwidths.

Specifications of the Keysight N8973A (Agilent) include:

  • Frequency Rage: 10 MHz - 3 GHz
  • Measurement Bandwidth (nominal): 4 MHz, 2 MHz, 1 MHz, 400 kHz, 200 kHz, 100 kHz
  • Frequency reference:
  •   Standard w/ Opt. 1D5
    Aging ± < 2 ppm4/year ± < 0.1 ppm/year
    Temperature Stability ± < 6 ppm ± < 0.01 ppm
    Settability ± < 0.5 ppm ± < 0.01 ppm
  • Tuning accuracy (dependent on bandwidth) (start, stop, center, marker)
    • 4 MHz measurement bandwidth (default on all models of noise figure analyzer)
    • Frequency Error
      10 MHz – 3.0 GHz ± < reference error + 100 kHz
         
    • < 4 MHz- measurement bandwidth (functionality not present in N8972A)
    • Frequency Error
      10 MHz – 3.0 GHz < ± reference error + 20 kHz
         
  • Noise Source ENR:
    •   4-7dB 12-17dB 20-22dB
      NF Measurement Range 0 to 20 dB 0 to 30 dB 0 to 35 dB
      NF Instrument Uncertainty ± < 0.05 dB ± < 0.05 dB ± < 0.1 dB
      Gain Measurement Range: -20 to +40 dB      
      Gain Instrument Uncertainty: ± < 0.17 dB      
  • Noise Figure of Instrument:
    • Frequency Noise Figure Noise figure over a limited temperature range of 23° C ± 3° C
      10 MHz to < 500 MHz < 4.9 dB + (0.0025 * freq in MHz) < 4.4 dB + (0.0025 * freq in MHz)
      500 MHz to < 2.3 GHz 500 MHz to < 2.3 GHz < 5.9 dB + (0.00135 * freq in MHz)
      2.3 GHz to 3.0 GHz 4.9 dB + (0.0015 * freq in MHz) < 2.9 dB + (0.0015 * freq in MHz)

Request A Quote

Rent      Buy      Repair
SUBMIT REQUEST
760-806-6600


Testimonial







Home   |   Rental Equipment   |   Equipment Sales   |   Repair & Calibration   |   Sell Your Equipment   |   About Us   |   Contact   |   Get A Quote   |   Blog    



YouTube Facebook Google Plus Paypal
Get the latest deals first. Subscribe to our newsletter.    SUBSCRIBE


© 2020 Axiom Test, All Rights Reserved
FAQ      Privacy      Terms      Site Map      Site Map XML     
© 2020 Axiom Test, All Rights Reserved
FAQ      Privacy      Terms      Site Map      Site Map XML

Axiom Test Equipment
2610 Commerce Way Vista, CA 92081
Phone: (760) 806-6600